D397-50-880 Issue C
Page 30 © Edwards Limited 2010. All rights reserved.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Operation
4.8.10 Software module 906 - Turbo Pump Power
This module can be used with any Turbo pump that is compatible with a TIC unit.
General Description
Monitors turbo power.
Output bytes (from Master)
None.
Input bytes (to Master)
No Alerts currently exist for this software module so the 3rd and 4th bytes will always be zeros.
Associated Parameter options
None.
4.8.11 Software module 907 - Turbo Normal
This module can be used with any Turbo pump that is compatible with a TIC unit.
General Description
Monitors whether turbo pump is at normal speed.
Output bytes (from Master)
None.
Input bytes (to Master)
No Alerts currently exist for these software modules so the 2nd and 3rd bytes will always be zero.
Associated Parameter options
None.
4.8.12 Software module 908 - Turbo Standby
This module can be used with any Turbo pump that is compatible with a TIC unit.
General Description
Set up a pump standby speed mode and monitors whether this has occurred.
Output bytes (from Master)
1st Form a 16-bit number, which is the motor power 10th's of Watts.
2nd
3rd Alert - refer to table in Appendix A2.
4th Priority - refer to table in Appendix A3.
1st Whether at Normal speed: 0 =No; 4 = Yes.
2nd Alert - refer to table in Appendix A2.
3rd Priority - refer to table in Appendix A3.
1st Standby command (1 = standby mode; 0= normal mode).