Configuration
[68]
CAUTION!
8.8
Components database add a new target
material
The target materials, available in the components database, are listed in the component
selection dialog of the configuration editor. If the required target material is not listed in
the selection dialog, follow the steps to add it to the components database.
Before adding a new target material to the components database, ensure that the required
material data is known, as the material has an influence on the accuracy of the measurement.
For additional target material data, go to the Emerson download portal. If the required
material data is not available on the portal, contact Product Support. If required, Product
Support will ask you to send in a sample of the target material to determine the material data.
The thickness of the sample should be not less than 5 mm and the diameter not less than 50
mm.
Procedure
1. Get the target material data from the download portal.
a. Enter http://reliabilitymobile.com/apps/registration/Account/Login.aspx into
the address bar of your internet browser.
b. Log in to the download portal.
Click Register if you do not already have an account. Follow the instructions.
c. Go to AMS EZ 1000 and download the target material data list to your computer.
If the required material is not on the list, contact Product Support.
2. Start Machine Studio, and go to File > Settings.
3. Select Components Database, and click Add.
The dialog for the selection of the components type opens.
4. Select Material, and click Next.
The dialog for entering the material data opens.
5. Enter a name for the material and the data from the already downloaded material
data list. Click Next.
The confirmation dialog opens.
6. Check your entries, and place a checkmark in the box Data is correct to confirm the
correctness of the entries. Click Finish.
Note
Once entered into the database, the material data cannot be changed, only deleted.