4 Wiring the channels
Note
To prevent conduit connectors from seizing in the threads of the conduit openings, apply
a conductive anti-galling compound to the threads, or wrap threads with PTFE tape two to
three layers deep.
Wrap the tape in the opposite direction that the male threads will turn when inserted into
the female conduit opening.
4.1 Available channels
Signal Channel A Channel B Channel C Channel D Channel E
Wiring
terminals
1 2 3 4 5 6 7 8 9 10
mA inputs
and outputs
mA output 1
(HART
®
)
mA output 2 mA output 3 mA input RS-485
Frequency
outputs
N/A Frequency output
2
(1)
Frequency output
1
Frequency output
2
(1)
N/A
Discrete
outputs
N/A Discrete output 1 Discrete output 2 Discrete output 3 N/A
Discrete
inputs
N/A N/A Discrete input 1 Discrete input 2 N/A
Frequency
inputs
N/A N/A N/A Frequency input N/A
(1) Frequency output 2 can be mapped to Channel B or D. For multiple frequency outputs, use Frequency 1 on Channel C
and Frequency 2 on either Channel B or D.
Installation Manual Wiring the channels
MMI-20027478 April 2022