4.6 Wire the discrete output
Use this section to wire the discrete output in explosion-proof, nonincendive, or
nonhazardous installations.
Important
Meter installation and wiring should be performed only by suitably-trained personnel.
4.6.1 Wire the discrete output (internally powered)
Use this procedure to wire the internally-powered discrete output for Channel B or C.
Procedure
Wire to the appropriate output terminal and pins.
Figure 4-16: Discrete output wiring (internally powered)
A. Discrete output
B. Channel B or C
C. See Figure 4-17 for output amplitude versus load resistance
D. Counter
Figure 4-17: Internally powered discrete output: output amplitude versus load
resistance [24 VDC (Nom) open circuit]
0
2
4
6
8
10
12
14
16
18
20
22
24
0 750 1500 2250 3000
A
B
A. Output amplitude (V)
B. Load resistor (Ω)
Wiring the channels Installation Manual
April 2022 MMI-20027478
34 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs