64 • RLDS I&O Manual 026-1309 Rev 8
For information about the Diagnostic Screen, refer to
Section 8.10., The Diagnostic Screen.
13.5.2. Gas Testing
To verify monitor operation, gas testing is
recommended. This ensures the integrity of the
sample tubing is still intact and the sensor is respond-
ing to gas. Gas testing can be done either by “bump
testing” or by performing a gas verification test.
A bump test is done by exposing each zone to
either a liquid or gaseous refrigerant. The test verifies
the sample path and checks that the sensor is
responding to gas. The refrigerant concentration is not
determined.
A gas verification test is done by exposing each zone
to a known concentration of gas and verifying that the
display readings are within 10% of the applied gas.
Gas verification kits for a variety of refrigerant types
are available which contain certified PPM
levels of test gas. For more information about gas
verification test, refer to Section 14.1.1., Testing with
Gas Verification Kit.
Figure 13-1 - Navigating to the Diagnostic Screen

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson RLDS and is the answer not in the manual?

Emerson RLDS Specifications

General IconGeneral
BrandEmerson
ModelRLDS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals