If a malfunction is found, refer to the table below to take the appropriate measures.
Never disassemble, modify or repair the device.
It can result in electrical shock.
Symptoms Causes and measures
The monitor and the power indicator
are not turned on.
• The power cord may not be properly connected.
Make sure to connect it securely.
• The fuse may be blown.
If so, replace it with the new one.
The fuse is blown when the power
switch is turned on.
• Contact your local distributor immediately.
The monitor display is suddenly
disappeared.
• The saving function may be activated.
Press any switch to deactivate the saving function.
The moving parts such as a joystick
are not moving appropriately.
• Do not move the part forcibly.
Contact your local distributor or service person.
It does not print out.
• Check if the papers are set.
Reload them if the papers are not set.
• The setting of Print REF/KRT may be set as OFF.
Change the setting.
The printer paper comes out but not
printing.
• The printer paper may be set in a wrong direction.
Set the paper properly.
The date setting is inaccurate.
• The battery in the device may be run out.
Keep the power on for 24 hours and recharge it.
Contact your local distributor immediately if the situation does not improve even after taking the measures
mentioned above.