7.1 Clean the instrument externally
CAUTION
Disconnect the power cord form the power outlet before cleaning
the instrument.
Maintenance interval
Clean the instrument externally when required. Do not allow spilled liquid to dry on the
instrument.
Required material
The following materials are required:
Cloth
Mild cleaning agent or 20% ethanol
Instruction
Follow the steps below to clean the instrument externally.
ActionStep
Turn off the instrument and disconnect the power cord from the wall socket.1
Wipe the surface with a damp cloth. Wipe off stains using a mild cleaning
agent or 20% ethanol. Wipe off any excess.
2
Remove the top and bottom tray for cleaning. Put them back once they are
cleaned.
3
Let the instrument dry completely before using it.4
ÄKTA go Operating Instructions 29360951 AC 91
7 Maintenance
7.1 Clean the instrument externally