Page #97 background image
7.4 Replace the main fuses
WARNING
Disconnect power. Always disconnect power from the instrument
before replacing fuses.
WARNING
For continued protection from fire hazard, replace only with same
type and rating of fuse.
WARNING
If a fuse requires repeated replacement, do not continue to use
the instrument. Contact an authorized service engineer.
Required tools
DimensionTool
2 to 3 mmFlat screwdriver
Procedure
Follow the instructions below to replace the fuses.
There are two identical fuses at the mains power inlet. Replace both fuses
even if only one is blown.
Note:
ActionStep
Disconnect the power cord from the power inlet.1
ÄKTA go Operating Instructions 29360951 AC 97
7 Maintenance
7.4 Replace the main fuses