EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ V3

GE LOGIQ V3 Service Manual

GE LOGIQ V3
170 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #50 background imageLoading...
Page #50 background image
GE LOGIQ V5/LOGIQ V3
D
IRECTION 5496012-100, REVISION 3BASIC SERVICE MANUAL
2 - 8 Section 2-2 - General Console Requirements
2-3-3 Minimal floor plan suggestion
CSI 8x10
Figure 2-1. Minimal floor plan, 2.5 m x 3 m (8 by 10 foot)
Scale:
Each square equals one square foot (app. 31 x 31 cm)
1. Door – at least 762 mm (30 inches)
2. Film Viewer
3. Counter Top, Sink with hot and cold
water and Supplies Storage
4. Linen Supply
5. Probes/Supplies
6. Examination Table – 1930 x 610 mm
(76 x 24 inches)
7. Footswitch
8. Stool
9. Ultrasound system
10. External Peripherals
11. Dedicated Power Outlet - Circuit
Breaker protected and easily
accessible
12. Network Interface
13. 457 mm (18 inches) distance of Ultrasound system
from wall or objects
14. GE Cabinet for Software and Manuals
4
1
3
11
6
13
8
10
5
7
9
12
14
2

Table of Contents

Other manuals for GE LOGIQ V3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V3 and is the answer not in the manual?

GE LOGIQ V3 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V3
CategoryMedical Equipment
LanguageEnglish

Related product manuals