Parameter Item List
All items that make up the operating parameters, codes, alarms, calibration references, voltage
levels and other information relative to operation or configuration of the Mercury Instruments
ERX Recorder are listed below. When connected and linked to a computer, all ERX Recorder
items can be displayed or changed in the Display/Item by Function of MasterLink32 software.
Item
Item Name & Description
061 Display Test
A test pattern generated by the microprocessor that gives a visual indication of the status
of LCD; pattern energizes all segments of LCD.
104 A/D FAULT
This item indicates if an alarm for the A/D Converter was generated. If the
microprocessor detects an A/D fault (Analog to Digital), an alarm is initiated and is
indicated by placing dots on
the LCD and “11111111” at Item 104. “00000000” at Item
104 indicates there is no A/D Converter Alarm. The alarm indicators will remain active
until manually cleared (by software). Unlike other instrument alarms, an A/D Fault is
usually a hardware failure that requires factory replacement of a component on the main
circuit board, or a field replacement of the main board.
Default = 00000000
118
Reference Number 1 000000000
119 Reference Number 2 00000000
127 Instrument Type 10
P3 Pressure Coefficients
150 P3 Pressure Coefficient 27 0.000000
151 P3 Pressure Coefficient 28 0.000000
152 P3 Pressure Coefficient 29 0.000000
153 P3 Pressure Coefficient 30 30.00000
154 P3 Pressure Coefficient 31 0.000000
155 P3 Pressure Coefficient 32 0.000000
156 P3 Interval High Pressu
re Time 00 00 00
157 P3 Interval High Pressure Date 01-01-04
158 P3 Interval Low Pressure Time 00 00 00
159 P3 Interval Low Pressure Date 01-01-04
173 AT Reset Log Trigger 1
Board Coefficients
178 P1 Board Coefficients A 0.0000
179 P1 Board Coefficients B 1.0000
180 P1 Board Coefficients C 0.0000
181 P1 Board Coefficients D 1.0000
265 Reserved MA 00000000
266 Reserved M
D 00000000
267 Reserved MC 00000000
Item Item Name & Description