System Commands
Copyright© Itech Electronic Co., Ltd. 15
Command syntax
SYST:LOC
Arguments
None
SYSTem:RWLock[:STATe]
This command is transmit through RS232 interface, the same function as
SYSTem:REMote, but the difference is: this command can lock “LOCAL”
button as well. When this command is executed, the “Local” button can’t
switch to local control mode.
Command syntax
SYST:RWL
Arguments
None
SYSTem:COMMunicate:GPIB:RDEVice:ADDRess
This command is used to set the device address of the GPIB communication.
Command syntax
SYSTem:COMMunicate:GPIB:RDEVice:ADDRess {<NR1>}
Arguments
1~30
Query syntax
SYSTem:COMMunicate:GPIB:RDEVice:ADDRess?
Returns
<NR1>
SYSTem:COMMunicate:LAN:CURRent:ADDRess
This command is used to set IP address of power supply. (LAN related
instructions only apply to IT6300C series instruments)
Command syntax
SYSTem:COMMunicate:LAN:CURRent:ADDRess <STR>
Arguments
<STR>
Unit
None