System Commands
Copyright© Itech Electronic Co., Ltd. 24
STATus:OPERation:ENABle
This command can set the parameter of operation event enable register.
Setting parameter can determine which bit value of operation event register is 1
and the bit will cause OPER of status byte register to be 1.
Command syntax
STATus: OPERation:ENABle <NRf>
Arguments
0~255
Default set
Refer to *PSC command
Example
STATus: OPERation:ENABle 128
Query syntax
STATus: OPERation:ENABle?
Returns
<NR1>
STATus:OPERation:INSTrument[:EVENt]?
This command can query the parameter of operation event register in the
channel. After executing this order, the register is reset.
Query syntax
STATus:OPERation:INSTrument[:EVENt]?
Arguments
None
STATus:OPERation:INSTrument:ENABle
This command can re-write the parameter of operation enable register in the
channel.
Command syntax
STATus:OPERation:INSTrument:ENABle
Arguments
Register parameter
Query syntax
STATus:OPERation:INSTrument:ENABle?

Table of Contents

Other manuals for ITech IT6322A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT6322A and is the answer not in the manual?

ITech IT6322A Specifications

General IconGeneral
BrandITech
ModelIT6322A
CategoryPower Supply
LanguageEnglish

Related product manuals