Measurement Commands
Copyright© Itech Electronic Co., Ltd. 43
Example
MEAS:CURR?
MEASure[:SCALar]:POWer[:DC]?
This command is used to read the output power of power supply.
Command syntax
MEASure[:SCALar]:POWer?
Arguments
None
Returns
<NR2>
Unit
W
Example
MEAS:POW?
MEASure[:SCALar]:CURRent:ALL[:DC]?
This command is used to return the current value of three channels at the same
time (compatible with the previous 6322).
Command syntax
MEASure[:SCALar]:CURRent:ALL[:DC]?
Arguments
None
MEASure[:SCALar][:VOLTage]:ALL[:DC]?
This command is used to return the voltage value of three channels at the same
time (compatible with the previous 6322).
Command syntax
MEASure[:SCALar][:VOLTage]:ALL[:DC]?
Arguments
None