Section
5: Leakage current and insulation resistance Model 2470 High Voltage SourceMeter Instrument
User's Manual
5-10 2470-900-01 Rev. A / May 2019
Set up the insulation resistance application using the front panel
To set up the application from the front panel:
1. Connect the device under test (DUT) to the rear panel of the 2470, as described in
Device
connections (on page 5-2).
2. Reset the 2470.
a. Press the MENU key.
b. Under System, select Info/Manage.
c. Select System Reset, and then select OK.
3. Press the TERMINALS FRONT/REAR switch to set the instrument to use the rear-panel
terminals (R is displayed to the left of the switch).
4. Press the HOME key.
5. Press the FUNCTION key.
6. Under Source Voltage and Measure, select Resistance.
7. Select the button next to Source (at the bottom of the screen).
8. Enter 700 V and select OK.
9. Press the MENU key.
10. Under Trigger, select Templates.
11. Set Templates to SimpleLoop.
12. Set Count to 10.
13. Set Delay to 0.1 seconds.
14. Press the HOME key.
15. Press the OUTPUT ON/OFF switch to enable the output.
16. Press the TRIGGER key to begin taking readings. The resistance measurements appear in the
m
easure area (top half) of the Home screen.
17. Press the OUTPUT ON/OFF switch to turn off the output when you are done maki
ng
m
easurements.