EasyManuals Logo
Home>Keithley>Test Equipment>263

Keithley 263 User Manual

Keithley 263
221 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #90 background imageLoading...
Page #90 background image
IEEE-488 PROGRAMMING
4.7.16
Y (Terminator)
Purpose
Format
Parameters
Default
Description
Use to select the ASCII terminator sequence that marks the end of the instrument’s
data string or status word.
Yn
YO CR LF
Yl LF CR
Y2 CR
Y3 LF
Y4 No Terminator
Upon power up, or after a DCL or SDC is received, the YO terminator (CR LF) will
be selected.
A terminator sequence can be programmed by sending the Y command followed
by an appropriate character. The default terminator sequence is the commonly us-
ed carriage return, line feed (CR LF) sequence (YO). Selecting the wrong terminator
for the controller could cause the bus to hang up.
The ASCII value of the CR character is 13, and the ASCII value of the LF character
is 10.
Programming OIJTPIJT TOE:.; c i ‘.:I)::’ J
! Terminate on LF CR.
Examples
,;,,JTP,jT 7fisi i i ‘.(,T::.::’ ?
! Restore default terminator
4-29

Table of Contents

Other manuals for Keithley 263

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 263 and is the answer not in the manual?

Keithley 263 Specifications

General IconGeneral
BrandKeithley
Model263
CategoryTest Equipment
LanguageEnglish

Related product manuals