S530/S540 KTE Linear Parametric Test Library (LPTLib) User's Manual Section 3:
LPTLib command reference
S530-900-01 Rev. E / September 2017 3-89
When multiple calls to the sweepX command are executed in the same test sequence, the smeasX,
sintgX, or savgX arrays are loaded sequentially. This appends the measurements from the second
sweepX call to the previous results. If the arrays are not dimensioned correctly, access violations
occur. The measurement table remains intact until the clrscn or devint command is executed.
Defining new test sequences using the smeasX, sintgX, or savgX commands adds commands to
the active measure list. The previous measures are still defined and used. The clrscn command is
used to eliminate the previous measures for the second sweep. Using the smeasX, sintgX, or
savgX command after a clrscn command causes the appropriate new measures to be defined and
used.
When the first sweep point is nonzero, it may be necessary to precharge the circuit so that the
sweepX command will return a stable value for the first measured point without penalizing remaining
points in the sweep. For example:
double ires[6];
conpin(SMU1, 10);
conpin(2, GND, 0);
forcev(SMU1, 5.0); /* Force 5 V to charge. */
delay(10); /* Wait for precharge. */
smeasi(SMU1, &ires); /* Set up the measurement. */
sweepv(SMU1, 5.0, 10.0, 5, /* Make the real measurement. */
2.5E-3);

Table of Contents

Other manuals for Keithley S530

Question and Answer IconNeed help?

Do you have a question about the Keithley S530 and is the answer not in the manual?

Keithley S530 Specifications

General IconGeneral
BrandKeithley
ModelS530
CategoryTest Equipment
LanguageEnglish

Summary

Safety Precautions

General Information

Introduction

Overview of the manual's structure and content.

Manual Contents

Details the structure and information presented in each section.

Linear Parametric Test Library (LPTLib)

Introduction

Introduces the Keithley LPTLib as a low-level software interface for parametric testers.

Measuring

Discusses the tester's ability to make measurements, including ordinary, integrated, and averaged.

Ranging

Discusses automatic range selection (autoranging) and its features like smart and sticky ranging.

Optimizing Test Sequences

Introduces methods to optimize the S530/S540 for faster operation.

LPTLib Command Reference

Conventions Used in this Manual

Explains conventions like case-sensitivity and parameter formatting.

Source Commands

Lists commands for controlling sources like forceX, limitX, pulseX.

Measure Commands

Lists commands for measurements like avgX, bmeasX, intgX.

Pulse Generator Commands

Lists commands for pulse generator units (PGU) like pgu_current_limit, pgu_delay.

Error Definitions

Result Values Indicating an Error

Table of error values returned as measured results and their meanings.

Error Messages

Lists error messages with numbers, descriptions, and remarks.

Related product manuals