5)
If
phase
difference
Is
negligible
and the
horizontal distance
is
short, press
the x10
MAG
switch
in
this
condition.
With
the x 10 MAG
switch pressed,
the
phase
difference
is 4.5
degrees
for
every
division.
Cursor
measurement
[CS-5230,
5260,
5270]
(1)
Carry
out
step
1 of
observation with
the
scale
on the CRT.
(2)
Adjust
the SWEEP
TIME/DIV
nad H.
VARI
controls
so
that
one
period
of the
waveform
is
5
div.
(3) Press
the CURSOR
MODE
switch
to
select
the
time
and
frequency
mode.
Align
the
AREF
cursor
and A
cursor
to the
corresponding points
of two
waveforms.
Read
the
measurement.
Figure
20-2
Measuring
Phase
Difference
(Cursor
measurement)
47
1 cycle
Reference signal
Comparison
signal
Phase
difference
Figure
20-1
Measuring
Phase
Difference
Measured
value
REF cursor
cursor