9.
FUSE
REPLACEMENT,
LINE
VOLTAGE
SELECTION
Before
conducting
the
following operation, turn
the power off and
unplug
the power
cord
from
the
wall socket.
Fuse
Replacement
If
the
fuse
blows,
the
osci1loscope
wi11 not
operate.
Locate
the
cause.
If the
oscilloscope
is not
defective,
remove the cap
from
the
fuse
holder
on the
rear
panel
with
a
standard
type
screwdriver,
then
replace
the
fuse with
a new
fuse.
(See
Figure
24.)
100
or 120 V : 1 A
(Time
lag)
220,
230 or 240 V : 630 mA
(Time
lag)
|
NOTE
1
If
any of the
following cases apply, please contact
your
dealer,
or our
distributor.
It
is not
clear
why the
fuse
has
blown,
or the
oscilloscope
seems to be
defective.
The
oscilloscope
uses
a
time
lag
type
fuse.
If a
time
lag
type
fuse
of the above-
shown
rating
is not
available.
Changing
Voltage
Select
ion
Remove
the
fuse
holder
and
voltage selector
from
the
rear
panel
with
a
standard
type
screwdriver.
Fit the 1ine
voltage selector
to match the
supply
voltage rat
ing
(T
mark).
If 100 or 120 V is to be changed
into
220, 230 or 240 V, the 1 A
fuse
should
be
replaced with
a 630 mA
fuse.
(See
Figure 24.
)
Fuse
Holder
Figure
24 Changing
Voltage
Requi rements
50

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kenwood CS-5200 Series and is the answer not in the manual?

Kenwood CS-5200 Series Specifications

General IconGeneral
BrandKenwood
ModelCS-5200 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals