46 AEJ-C/AES-C/PLJ-C-BA-e-1621
9.3.6 <P2.2.2 > AUTOTARE / <P2.2.3.2> AUTO THRESHOLD
The automatic tare function automatically saves the first stable weighing value as
tare value.
The parameter <P2.2.3.2> AUTO THRESHOLD is used to define which minimum
weight must be placed before automatic taring takes place.
Further automatic taring ill only take place when the display drops below the set
value for the auto threshold.