AEJ-C/AES-C/PLJ-C-BA-e-1621 165
29.6.15 Query lower tolerance unit
Command: ODH CR LF
Response:
1 2 3 4 - 12 13 14 - 16 17 18 19
D H
˽
Tare
˽
Unit
˽
CR LF
Weight: - 9 characters right aligned
Unit: - 3 characters left aligned
29.6.16 Query upper tolerance unit
Command: OUH CR LF
Response:
1 2 3 4 - 12 13 14 - 16 17 18 19
U H
˽
Tare
˽
Unit
˽
CR LF
Weight: - 9 characters right aligned
Unit: - 3 characters left aligned
29.6.17 Enter reference single weight
Command: SM˽XXXXX CR LF
Possible replies:
SM˽OK CR LF
Command executed
SM˽I CR LF
Command valid but cannot be executed at the moment (e. g.
incorrect operating mode)
ES CR LF
Syntax error (incorrect weight format)
29.6.18 Enter target weight (dosing)
Command: TV˽XXXXX CR LF
Possible replies:
TV˽OK CR LF
Command executed
TV˽I CR LF
Command valid but cannot be executed at the moment (e. g.
incorrect operating mode)
ES CR LF
Syntax error (incorrect weight format)
DH˽MASA CR LF
Command executed
UH˽MASA CR LF
Command executed