164 AEJ-C/AES-C/PLJ-C-BA-e-1621
29.6.11 Start continuous output in current weighing unit
Command: CU1˽CR LF
Possible replies:
CU1˽I CR LF
Command valid but cannot be executed at the moment
CU1˽A CR LF
Command accepted, command executing
Weighing values are shown in current weighing unit
S U I
˽
Signs Weight
˽
Unit CR LF
29.6.12 Stop continuous output in current weighing unit
Command: CU0˽CR LF
Possible replies:
CU0˽I CR LF
Command valid but cannot be executed at the moment
CU0˽A CR LF
Command accepted, command executing
29.6.13 Enter lower tolerance limit
Command: DH˽XXXXX CR LF
Possible replies:
DH˽OK CR LF
Command executed
ES CR LF
Syntax error (incorrect weight format)
29.6.14 Enter upper tolerance limit
Command: UH˽XXXXX CR LF
Possible replies:
UH˽OK CR LF
Command executed
ES CR LF
Syntax error (incorrect weight format)