EasyManuals Logo
Home>KERN>Scales>AEJ-C Series

KERN AEJ-C Series User Manual

KERN AEJ-C Series
Go to English
513 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #280 background imageLoading...
Page #280 background image
108 AEJ-C/AES-C/PLJ-C-BA-e-1621
20.2 Perform percentage weighing
Preparation
For assigning two function keys of your choice with the functions <ENTER
REFERENCE> and <DEFINE REFERENCE WEIGHT> see chap. 20.1 or chap.
9.3.10.
20.2.1 Determination of the reference weight by weighing
For selecting application <DEVIATION> see chap.16.1. The reference weight
used will be shown in the informational line of text. At the first start-up as value
0.0000g.
Press the function key assigned to the <DEFINE REFERENCE WEIGHT)
function (display example F1)
Put down a reference weight equalling 100 %, wait until the stability sign is
shown on the weighing balance screen and confirm by pressing the PRINT-
key.
The weight will be adopted as reference (100%) and shown in the
informational line of text.
Tare, as required and place load on pan. The weight of the item to be weighed
is displayed in percent, referring to the reference weight.
The reference weight will remain saved until a new reference is defined.
Return to weighing mode, see chap. 17.2.1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN AEJ-C Series and is the answer not in the manual?

KERN AEJ-C Series Specifications

General IconGeneral
BrandKERN
ModelAEJ-C Series
CategoryScales
LanguageEnglish

Related product manuals