78
Version 2.0, Revision I
Troubleshooting
7
Display
code
Source of error Troubleshooting solution
A print job was attempted with a storage
fluid cartridge inserted.
Remove storage cartridge. Insert ink
cartridge and press LOADED to confirm
( p. 41 – 4.9 Exchanging the transport
cartridge for an ink cartridge).
Power supply unit does not achieve
standard operating voltage.
Call Leica Technical Service.
Firmware only partially loaded or not
loaded at all.
Flash memory defective.
Call Leica Technical Service.
Wrong firmware. Call Leica Technical Service.
7.4 Changing the flash bulb
Code 13 appears on the printer display when the flash bulb has reached its maximum service life.
When code 13 is issued, the bulb has to be changed.
For details on how to insert/replace the flash bulb, please refer to ( p. 32 – 4.6 Installing/exchanging
the flash bulb).
Warning
Before replacing the flash bulb, switch off the printer and unplug the instrument from power supply.
10 s
1. After changing the bulb, switch the printer back on.
2. Then go to offline mode and hold down the ERROR key for 10 seconds; code "13" disappears
from the display.