5-2
5.3 Basic Disassembly
5.3.1 Disconnecting the Base
NOTE
z Be sure to disassemble the base first before proceeding with other parts.
z Be sure to place the monitor face up when disassembling it. If lay the monitor face
down, be sure the surface is non-abrasive and static-free. Otherwise, the screen or
the knob may be scratched or damaged.
1. The hook that prevents the power plug off is located beside the AC port in the rear case
of the monitor. Remove this hook and then place the monitor face up and unscrew the
four M4×12 screws, as shown in the figure below.