5-10
5.4.5 Removing the Inverter of T6 and T8
1. Unscrew the six M3×6 screws and remove the inverter shield.
2. Unplug the cables respectively from the button board and the LCD to the inverter. Then,
unscrew the two M3×6 screws to remove the inverter.
NOTE
z Release the clip (if any) before disconnecting the cable between the LCD and the
inverter.
z When reassembling, check whether the two heat-conductive insulators are in place.
Heat-conductive insulator