3-2
3.1.3 Preventative Maintenance
Below are preventative maintenance tests which need to be performed on the monitor. See the
following sections for detailed maintenance procedures.
Visual inspection
NIBP test and calibration
Microsteam and Sidestram CO
2
test and calibration
AG test and calibration
3.1.4 Recommended Frequency
Check/Maintenance Item Frequency
Preventative Maintenance Tests
Visual inspection 1. When first installed or reinstalled.
Pressure check
Leakage test
NIBP test
Calibration
Leakage test
Performance test
Sidestream and
Microstream
CO
2
tests
Calibration
Performance test
AG tests
Calibration
1. If the user suspects that the measurement is
incorrect.
2. Following any repairs or replacement of relevant
module.
3.At least once a year.
4. AG leakage test should be performed before AG
measurement.
Performance Tests
Performance test
ECG test
Calibration
Resp performance test
SpO
2
test
Pressure check
Leakage test
NIBP test
Calibration
Temp test
IBP test
Performance test
1. If the user suspects that the measurement is
incorrect.
2. Following any repairs or replacement of relevant
module.
3. At least once every two years. At least once a year
is recommended for NIBP, CO
2,
NMT and AG.
4. AG leakage test should be performed before AG
measurement.

Table of Contents

Other manuals for Mindray BeneView T6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneView T6 and is the answer not in the manual?

Mindray BeneView T6 Specifications

General IconGeneral
BrandMindray
ModelBeneView T6
CategoryMedical Equipment
LanguageEnglish

Related product manuals