4.2.4. Submenu „Display settings“- „* LCD“
Under this menu point the contrast and displayed units can be defined.
„CONTRA“
For an optimal display reading the contrast can be adapted individually.
LC-Display contrast: 0....9
„UNIT“
The LabSwift-aw can display the measured values (results) in various units.
„UNIT H“
Humidity / AW: aw-value or relative humidity in % RH
„UNIT T“
Temperature: °C or °F
„EXIT“
Quit the submenu „Display settings“
4.2.5. Submenu „SD-card“ - „* SD LOG“
„S NUMB“
Free number setting (format 0000-9999) as start number and for generation of
a data number in order to match the data to the correspondent measurement.
After each measurement this number is increased by 1.
„S INT“
The data storage interval on the SD-card can be selected in minutes and
seconds (0:00). For standard measurements and analysis with the Novalog
MC software a 10 seconds (0:10) interval is recommended. For data import to
Excel, longer interval periods might be wiser in order to reduce the data
quantity.
„EXIT“
Quits the submenu „SD-card“
4.3. Reset to factory settings
All settings are reset to factory settings except the stored calibration values on
the CM-2 measurement cells. This function can be performed as following:
For this purpose keep the button [up] whilst switching on the instrument. The
message „FA SET?“ appears on the display. Select „YES“ and the confirmation
message „Done“ appears and the device is switched off.
If you answer the question with „NO“, no data are reset and the device is
switched off „OFF“.