Chapter 3 Cleaning, Disinfection and Sterilization Procedures
21
EVIS EXERA GIF/CF/PCF TYPE 160 Series REPROCESSING MANUAL
Inspection of the equipment
For inspection of other equipment than that mentioned below, refer to the
instruction manual for the equipment being used.
The cleaning brushes are consumable items. Should the
slightest irregularity be suspected, use a spare instead.
Using a defective brush may cause equipment damage.
Inspection of the water-resistant cap (MH-553)
1. Confirm that the inside of the water-resistant cap is dry and free from debris
(see Figure 3.2).
2. Confirm that the seal inside the water-resistant cap is free from scratches,
flaws and debris.
3. Check the venting connector for looseness.
Inspection of the water-resistant cap (MAJ-583 for
CF-Q160ZL/I only)
1. Confirm that the inside of the water-resistant cap is dry and free from debris
(see Figure 3.3).
2. Confirm that the seal of the water-resistant cap is free from scratches, flaws
and debris.
3. Confirm that the thread of the water-resistant cap is free from deformation
and other flaws.
Inspection of the water-resistant cap (MAJ-942 for
CF-Q160DL/I only)
1. Confirm that the inside of the water-resistant cap is dry and free from debris
(see Figure 3.4).
2. Confirm that the seal of the water-resistant cap is free from scratches, flaws
and debris.
3. Confirm that the thread of the water-resistant cap is free from deformation
and other flaws.