EasyManuals Logo

Olympus GIF XTQ160 User Manual

Olympus GIF XTQ160
74 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #33 background imageLoading...
Page #33 background image
Chapter 3 Cleaning, Disinfection and Sterilization Procedures
29
EVIS EXERA GIF/CF/PCF TYPE 160 Series REPROCESSING MANUAL
Flush detergent solution and air into the auxiliary water
channel (for endoscopes with auxiliary water feeding only)
Do not cap the auxiliary water inlet during reprocessing.
If the auxiliary water tube is not connected to the auxiliary
water inlet, uncap the auxiliary water inlet and connect the
auxiliary water tube (see Figure 3.15).
Prior to reprocessing endoscopes equipped with auxiliary
water feeding, attach the auxiliary water inlet cap to the
auxiliary water inlet and open the cap (see Figure 3.15).
1. Immerse the distal end of the insertion tube in the water.
2. Using the 30 cm
3
(30 ml) syringe, slowly flush detergent solution through
the auxiliary water channel several times until no bubbles exit the distal end.
3. Using the 30 cm
3
(30 ml) syringe, slowly flush water through the auxiliary
water channel several times.
4. Using the 30 cm
3
(30 ml) syringe, slowly flush air through the auxiliary water
channel several times until a steady stream of bubbles exits the distal end.
Figure 3.15
Luer port
Auxiliary water inlet
Endoscope connector
Auxiliary water inlet cap
Clean
water
Syringe
Auxiliary water tube

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Olympus GIF XTQ160 and is the answer not in the manual?

Olympus GIF XTQ160 Specifications

General IconGeneral
BrandOlympus
ModelGIF XTQ160
CategoryMedical Equipment
LanguageEnglish

Related product manuals