48
Chapter 3 Cleaning, Disinfection and Sterilization Procedures
EVIS EXERA GIF/CF/PCF TYPE 160 Series REPROCESSING MANUAL
10. Detach the channel plug and injection tube from the endoscope.
11. Detach the auxiliary water tube from the endoscope (for endoscopes with
auxiliary water feeding only).
Figure 3.30
Dry external surfaces
1.
Using a clean, lint-free cloth, thoroughly wipe and dry the external surfaces
of the endoscope, channel plug and injection tube.
2. Inspect the endoscope for residual debris. Should debris remain, repeat the
procedures given in this section.