Chapter 3 Cleaning, Disinfection and Sterilization Procedures
EVIS EXERA GIF/CF/PCF TYPE 160 Series REPROCESSING MANUAL
3.9 Cleaning, disinfection and sterilization
procedures for reusable parts and reprocessing
equipment
This section includes the cleaning, disinfection and sterilization procedures for
the reusable parts and reprocessing equipment listed below. For all other parts,
refer to their respective instruction manuals. For compatible reprocessing
methods, refer to Section 2.1, “Compatibility summary”.
Equipment needed
Reusable parts which are normally reprocessed with the
endoscope
Personal protective equipment
Soft brush
Large basin
Detergent solution
Clean water (for non-sterile water rinse)
Clean, lint-free cloths
Disinfectant solution
Sterile water (for sterile water rinse)
Sterile, lint-free cloths
Small container
70% ethyl or isopropyl alcohol
•30cm
3
(30 ml) syringe
Channel plug (MH-944)
Injection tube (MH-946)
Water-resistant cap (see Table 1.1)