Chapter 3 Cleaning, Disinfection and Sterilization Procedures
EVIS EXERA GIF/CF/PCF TYPE 160 Series REPROCESSING MANUAL
Flushing detergent solution into the auxiliary water channel
(for endoscopes with auxiliary water feeding only)
Using the auxiliary water tube and 30 cm
3
(30 ml) syringe, inject detergent
solution into the auxiliary water channel.
1. Attach the auxiliary water tube to the auxiliary water inlet (see Figures 3.28
and 3.29).
2. Using the 30 cm
3
(30 ml) syringe, inject 90 cm
3
(90 ml) of detergent
solution.
3. Disconnect the auxiliary water tube from the endoscope and immerse it in
detergent solution.
Figure 3.28
Figure 3.29