46
Chapter 3 Cleaning, Disinfection and Sterilization Procedures
EVIS EXERA GIF/CF/PCF TYPE 160 Series REPROCESSING MANUAL
Flushing detergent solution into the auxiliary water channel
(for endoscopes with auxiliary water feeding only)
Using the auxiliary water tube and 30 cm
3
(30 ml) syringe, inject detergent
solution into the auxiliary water channel.
1. Attach the auxiliary water tube to the auxiliary water inlet (see Figures 3.28
and 3.29).
2. Using the 30 cm
3
(30 ml) syringe, inject 90 cm
3
(90 ml) of detergent
solution.
3. Disconnect the auxiliary water tube from the endoscope and immerse it in
detergent solution.
Figure 3.28
Figure 3.29
Syringe
Auxiliary water tube
Auxiliary water channel
Universal cord
Endoscope connector
Auxiliary water inlet cap
Auxiliary water inlet
Auxiliary water tube
Syringe