Chapter 3 Cleaning, Disinfection and Sterilization Procedures
47
EVIS EXERA GIF/CF/PCF TYPE 160 Series REPROCESSING MANUAL
Soaking the endoscope and all equipment in detergent
solution
1.
Using a clean, lint-free cloth, wipe all debris from the endoscope’s external
surfaces while the endoscope is immersed in the detergent solution.
2. Cover the basin with a tight fitting lid to minimize the release of detergent
vapors.
3. Soak the endoscope and all equipment for the amount of time and at the
temperature recommended by the detergent manufacturer.
Removing and rinsing the endoscope and all equipment
1.
Remove the endoscope and all equipment from the detergent solution and
place them in clean water.
2. Gently agitate them to thoroughly rinse.
Removing detergent solution from all channels
1.
Connect the channel plug and the injection tube to the endoscope. Place the
suction port in clean water.
2. Attach the 30 cm
3
(30 ml) syringe to the injection tube’s air/water channel
port and inject 90 cm
3
(90 ml) of clean water into the air/water channel (see
Figure 3.27).
3. Attach the 30 cm
3
(30 ml) syringe to the suction channel port and inject
90 cm
3
(90 ml) of clean water into the suction channel (see Figure 3.30).
4. Attach the auxiliary water tube to the auxiliary water inlet (for endoscopes
with auxiliary water feeding only).
5. Attach the 30 cm
3
(30 ml) syringe to the luer port of the auxiliary water tube
and inject 90 cm
3
(90 ml) of clean water to the auxiliary water channel (for
endoscopes with auxiliary water feeding only).
6. Remove the endoscope, together with all equipment, from the water.
7. Attach the 30 cm
3
(30 ml) syringe to the suction channel port and flush the
suction channel with 90 cm
3
(90 ml) of air (see Figure 3.30).
8. Attach the 30 cm
3
(30 ml) syringe to the injection tube’s air/water channel
port and flush the air/water channel with 90 cm
3
(90 ml) of air (see Figure
3.27).
9. Attach the 30 cm
3
(30 ml) syringe to the luer port of the auxiliary water tube
and flush the auxiliary water channel with 90 cm
3
(90 ml) of air (for
endoscopes with auxiliary water feeding only).

Table of Contents