23
English
English
Technical information
Single electric breast pump Double electric breast pump
Vacuum level
Stimulation: -60 to -200 mbar (-45 to -150 mmHg)
Expression: -60 to -360 mbar (-45 to -270 mmHg)
Adjustable in steps of 20 mbar
Intermittent vacuum
Cycle speed
Stimulation: 105-120 cycles/min, depending on vacuum level
Expression: 53-85 cycles/min, depending on vacuum level
Motor unit rated input 5 V direct current / 1.1 A 5 V direct current / 1.8 A
Mains adapter rated input
External mains adapter 100-240 V alternating current
50/60 Hz
400 mA
Mains adapter rated output 5 V direct current / 1.1 A 5 V direct current / 1.8 A
Adapter type number
S009AHz050yyyy The letters "yyyy" represent the output current from
0100 (1000mA) to 0180 (1800mA), increments in steps of 100mA. The
letter "z" denotes type of plug, it can be: U, V, B, S, C, A, D, K, T, I, E, F.
Power port
Micro-USB
Operating conditions
Temperature: 5°C to 40°C
Relative humidity: 15 to 90% (non-condensing)
Storage and transportation
conditions
Temperature: -25°C to 70°C
Relative humidity: up to 90% (non-condensing)
Operating and storage
pressure
700-1060 hPa (<3000m altitude)
Net weight 230g 310g
External dimensions
Motor unit: 145 mm x 95 mm x 45 mm (LxWxH)
Device classification
Applicable classifications:
Degree of protection against electric shock;
- Class II ME equipment
- Type BF applied part
The breast pump and the adapter are specified as medical electrical system
according to IEC 60601-1.
Materials
ABS, silicone (motor unit)
Polypropylene, silicone (other parts)
Service life 2 years
Operational life 500 hours
Mode of operation Continuous operation
Ingress protection IP22 (motor unit), IP20 (adapter)