- 97 -
N
T
space
stability
marker
zero marker
range
marker
digit marker
space
mass
space
mass unit
space
tare
space
tare unit
space
digits
CR
LF
- [space] if stable, [?] if unstable
- [space] for any value but zero, [Z]
for zero value
- marker informing on range within
which mass value is comprised:
[space] I range balance, [2] II
balance, [3] III balance
[zero] no digit, [1] one digit marker, [2]
two digits marker, [3] three digits
marker, [4] four digits marker, [5] five
10 characters for net mass given in
adjustment unit (with dot and a
floating-point “-“ sign )
3 characters with left justification
9 characters with dot - right
justification (when floating-
gets turned off auto
matically then
zero value is sent)
3 characters with left justification
Number specifying quantity of hidden
digits: [space] when no digits are
hidden, [1] when 1 digit is hidden
- command sent from a computer
N T _ ? _ _ 0 _ _ _ _ _ - 5 . 1 1 3 _ g _ _ _ _ _ _ _ 0 . 0 0 0 _ g _ _ _0 CR LF – command carried
out, response: mass value and remaining data
16. ERROR MESSAGES
-Err2- Value beyond zero range
-Err3- Value beyond tare range
-Err8- Taring / Zeroing operation time exceeded
-NULL- Zero value from converter
-FULL- Measurement range exceeded
-LH- Start mass error
17. ADDITIONAL ECQUIPMENT
Anti-vibration table
It is a very stable basis which eliminates all kinds of vibrations and ground shakes.
The inner part of the table holds a marble plate, which is a foundation for balance
positioning.