Remote Control
R&S
®
FPC
214User Manual 1178.4130.02 ─ 07
SYSTem:MEMory:USED: Queries how much memory is currently used.
Return values:
<MemorySize> <numeric value>
Example: //Query free RAM
SYST:MEM:FREE?
Usage: Query only
SYSTem:TIME <Hour>, <Minutes>, <Seconds>
This command defines the time of the internal clock.
Parameters:
<Hour> <numeric value> (integer)
Range: 0 to 23
<Minutes> <numeric value> (integer)
Range: 0 to 59
<Seconds> <numeric value> (integer)
Range: 0 to 59
Example: //Change system time
SYST:TIME 12,30,30
Manual operation: See "Date and Time" on page 356
SYSTem:TZONee <Hour>, <Minutes>
This command defines a shift of the system time to select another time zone.
Parameters:
<Hour> <numeric value> (integer)
Range: 0 to 23
*RST: 0
<Minutes> <numeric value> (integer)
Range: 0 to 59
*RST: 0
Example: //Shift the time one hour ahead
SYST:TZON 01,00
Manual operation: See "Date and Time" on page 356
UNIT<n>:LENGth <Unit>
This command selects the unit for distances.
Suffix:
<n>
.
irrelevant
System Configuration