Remote Control
R&S
®
FPC
337User Manual 1178.4130.02 ─ 07
[SENSe:]SWEep:POINts <SweepPoints>
This command defines the number of measurement points.
Parameters:
<SweepPoints> <numeric value> (integer only)
Range: 101 to 2501
*RST: 201
Example: //Define number of measurement points
SWE:POIN 1001
Manual operation: See "Number of Points" on page 401
19.16.6
Trace
Commands to configure traces described elsewhere.
Trace mode:
DISPlay<n>[:WINDow]:TRACe<t>:MODE
Note: Only "Clear/Write" and "Average" are supported by the VNA application.
Memory trace:
CALCulate<n>:MATH<t>:COPY:MEMory
DISPlay<n>[:WINDow]:TRACe<t>:MEMory[:STATe]
Data format of trace export:
FORMat[:DATA]
TRACe[:DATA]?.............................................................................................................337
TRACe[:DATA]?
This command queries the trace data (measurement results in graphical result dis-
plays).
Return values:
<Result> The number of values depends on the number of measurement
points.
The unit depends on the measurement and the measurement
format.
Example: //Query measurement results
TRAC?
Usage: Query only
19.16.7 Marker
Commands to configure markers described elsewhere.
Marker state (on / off):
CALCulate<n>:MARKer<m>[:STATe]
CALCulate<n>:MARKer<m>:AOFF
Remote Commands of the VNA Application