Remote Control
R&S
®
FPC
336User Manual 1178.4130.02 ─ 07
Parameters:
<State> ON | 1
Frequency settings are coupled. When you change the fre-
quency or span in the DTF measurement, they also change in
the other VNA measurements (reflection, transmission and cable
loss).
OFF | 0
Frequency settings of the DTF measurement are independent.
When you change the frequency or span in the DTF measure-
ment, they do not change in the other VNA measurements
(reflection, transmission and cable loss).
*RST: ON
Example: //Decouple frequency and span
FREQ:SETT:COUP:ENAB OFF
Manual operation: See "Coupled Span" on page 399
[SENSe:]FREQuency:SPAN:AUTO
<State>
This command turns the automatic calculation of the ideal span on and off.
Parameters:
<State> ON | OFF | 1 | 0
*RST: OFF | 0
Example: //Turn on automatic span
FREQ:SPAN:AUTO ON
Manual operation: See "Manual Span" on page 369
See "Manual Span" on page 399
19.16.4 Bandwidth
Commands to configure the bandwidth described elsewhere.
●
Resolution bandwidth:
[SENSe:]BANDwidth[:RESolution]
[SENSe:]BANDwidth[:RESolution]:AUTO
19.16.5 Sweep
Commands to configure the sweep described elsewhere.
●
Sweep mode:
INITiate:CONTinuous
●
Initiate measurement:
INITiate[:IMMediate]
[SENSe:]SWEep:POINts.................................................................................................337
Remote Commands of the VNA Application