Remote Control
R&S
®
FPC
340User Manual 1178.4130.02 ─ 07
DISPlay<n>:IMPedance:REFerence:POSition <Impedance>
This command defines the reference impedance for the Smith chart measurement for-
mat. The impedance can be between 1mΩ and 10 kΩ.
Prerequisites for this command
●
Select reflection measurement (MEASurement<n>:MODE).
●
Select Smith chart format (MEASurement<n>:FORMat).
Suffix:
<n>
.
irrelevant
Parameters:
<Impedance> <numeric value>
Range: 1 mOhm to 10 kOhm
*RST: 50 Ohm
Default unit: Ohm
Example: //Define reference impedance
MEAS:MODE REFL
MEAS:FORM SMIT
DISP:IMP:REF:POS
Manual operation: See "Marker Mode" on page 403
19.16.8 Limit Lines
Commands to configure limit lines described elsewhere.
●
Display lines:
CALCulate<n>:DLINe
CALCulate<n>:DLINe:STATe
●
Limit line management:
CALCulate<n>:LIMit<k>:COMMent?
CALCulate<n>:LIMit<k>:DEFine
CALCulate<n>:LIMit<k>:DELete
CALCulate<n>:LIMit<k>:UNIT:X?
CALCulate<n>:LIMit<k>:UNIT[:Y]?
●
Limit checks:
CALCulate<n>:LIMit<k>:BEEP[:STATe]
CALCulate<n>:LIMit<k>:FAIL?
CALCulate<n>:LIMit<k>:LOWer:SELect
CALCulate<n>:LIMit<k>:STATe
CALCulate<n>:LIMit<k>:UPPer:SELect
●
Threshold lines:
CALCulate<n>:LIMit<k>:LOWer:THReshold
CALCulate<n>:LIMit<k>:UPPer:THReshold
Remote Commands of the VNA Application