Remote Control
R&S
®
FPC
339User Manual 1178.4130.02 ─ 07
NADMittance standardized admittance in complex format (real +
imaginary)
NIMPedance standardized impedance in complex format (real +
imaginary)
NORMal normal marker
RPDB reflection coefficient in complex format (magnitude
(dB) + phase)
RPL reflection coefficient in complex format (magnitude
(lin) + phase
RSCalar reflection coefficient in complex format (real + imagi-
nary)
CALCulate<n>:MARKer<m>:Y?
This command queries the measurement results at the marker position.
If necessary, the corresponding marker is activated first.
To get a valid result, you have to perform a complete sweep with synchronization to the
sweep end between activating the delta marker and reading out the result. This is only
possible in single sweep mode.
Suffix:
<n>
.
irrelevant
<m> Marker
Return values:
<Position> <numeric value>
The unit of the return value depends on the measurement mode
and format you have selected.
In the Smith chart, the return value also depends on the marker
format you have selected with CALCulate<n>:MARKer<m>:
MODE.
Example: //Turn on single sweep mode and marker 2
INIT:CONT OFF
CALC:MARK2 ON
//Start a sweep and wait for the end
INIT;*WAI
//Query the position of marker 2
CALC:MARK2:X?
CALC:MARK2:Y?
Usage: Query only
Manual operation: See "Marker <x>" on page 403
Remote Commands of the VNA Application