EasyManuals Logo
Home>R&S>Measuring Instruments>FPC Series

R&S FPC Series User Manual

R&S FPC Series
419 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #327 background imageLoading...
Page #327 background image
Remote Control
R&S
®
FPC
324User Manual 1178.4130.02 ─ 07
19.16.1.1 Measurements and Result Display
MEASurement<n>:MODE...............................................................................................324
MEASurement<n>:FORMat............................................................................................ 324
MEASurement<n>:MODE <Measurement>
This command selects the measurement.
Suffix:
<n>
.
irrelevant
Parameters:
<Measurement> DTFault
Select the distance to fault measurement.
LOSS
Selects the 1-port cable loss measurement.
REFLection
Selects the reflection S11 measurement.
TRANsmission
Selects the transmission S21 measurement.
*RST: REFLection
Example: //Select measurement
MEAS:MODE LOSS
Manual operation: See "Reflection S11" on page 396
See "Transmission S21" on page 396
See "1-Port Cable Loss" on page 396
See "Distance To Fault" on page 396
MEASurement<n>:FORMat <Format>
This command select the measurement format,
Suffix:
<n>
.
irrelevant
Remote Commands of the VNA Application

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FPC Series and is the answer not in the manual?

R&S FPC Series Specifications

General IconGeneral
BrandR&S
ModelFPC Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals