DISPlay<n>:LOSS:REFerence <RefValue>
This command defines the reference value for the cable loss measurement format.
Prerequisites for this command
●
Select cable loss measurement (MEASurement<n>:MODE).
●
Select cable loss format (MEASurement<n>:FORMat).
Suffix:
<n>
.
irrelevant
Parameters:
<RefValue> <numeric value>
*RST: 0
Default unit: dB
Example: //Define reference value for cable loss
MEAS:MODE LOSS
MEAS:FORM LOSS
DISP:LOSS:REF 50
Manual operation: See "Reference Level" on page 399
DISPlay<n>:LOSS:REFerence:POSition <RefPosition>
This command defines the position of the reference value for the cable loss measure-
ment format.
Each step shifts the reference position one grid line up or down.
Prerequisites for this command
●
Select cable loss measurement (MEASurement<n>:MODE).
●
Select cable loss format (MEASurement<n>:FORMat).
Suffix:
<n>
.
irrelevant
Parameters:
<RefPosition> <numeric value> (integer only)
*RST: 8
Example: //Define 5th grid line as the reference position for cable loss
MEAS:MODE LOSS
MEAS:FORM LOSS
DISP:LOSS:REF:POS 5
Manual operation: See "Range / Ref Position" on page 399
DISPlay<n>:LOSS:Y:SCALe <DisplayRange>
This command defines the range of the y-axis for the cable loss measurement format.