EasyManuals Logo
Home>Rigol>Test Equipment>DS1104Z-S Plus

Rigol DS1104Z-S Plus Programming Guide

Rigol DS1104Z-S Plus
260 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #188 background imageLoading...
Page #188 background image
RIGOL Chapter 2 Command System
2-172 MSO1000Z/DS1000Z Programming Guide
:TRIGger:SLOPe:BLEVel
Syntax
:TRIGger:SLOPe:BLEVel <level>
:TRIGger:SLOPe:BLEVel?
Description
Set or query the lower limit of the trigger level in slope trigger. The unit is the same as the
current amplitude unit.
Parameter
Name
Type
Range
Default
<level>
Real
(-5 x VerticalScale - OFFSet) to
(5 x VerticalScale - OFFSet)
0
Explanation
For VerticalScale, refer to :CHANnel<n>:SCALe. For OFFSet, refer
to :CHANnel<n>:OFFSet.
Return
Format
The query returns the lower limit of the trigger level in scientific notation.
Example
:TRIGger:SLOPe:BLEVel 0.16 /*Set the lower limit of the trigger level to 160mV*/
:TRIGger:SLOPe:BLEVel? /*The query returns 1.600000e-01*/
Related
Command
:TRIGger:SLOPe:ALEVel

Table of Contents