EasyManuals Logo
Home>Rigol>Test Equipment>DS1104Z-S Plus

Rigol DS1104Z-S Plus Programming Guide

Rigol DS1104Z-S Plus
260 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #241 background imageLoading...
Page #241 background image
Chapter 2 Command System RIGOL
MSO1000Z/DS1000Z Programming Guide 2-225
:WAVeform:STOP
Syntax
:WAVeform:STOP <stop>
:WAVeform:STOP?
Description
Set or query the stop point of waveform data reading.
Parameter
Name
Type
Range
Default
<stop>
Integer
NORMal: 1 to 1200
MAX: 1 to the number of effective point
currently on the screen
RAW: 1 to the current maximum memory depth
1200
Explanation
When reading the waveform data in the internal memory, the actual settable ranges of the
start point and stop point of a reading operation are related to the memory depth of the
oscilloscope and the return format of the waveform data currently selected; for the
details, refer to the explanation in the :WAVeform:DATA? command.
Return
Format
The query returns an integer.
Example
:WAVeform:STOP 500 /*Set the stop point to 500*/
:WAVeform:STOP? /*The query returns 500*/
Related
Commands
:ACQuire:MDEPth
:WAVeform:MODE
:WAVeform:FORMat
:WAVeform:STARt

Table of Contents

Other manuals for Rigol DS1104Z-S Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1104Z-S Plus and is the answer not in the manual?

Rigol DS1104Z-S Plus Specifications

General IconGeneral
BrandRigol
ModelDS1104Z-S Plus
CategoryTest Equipment
LanguageEnglish

Related product manuals