RIGOL Chapter 2 Command System
2-40 MSO1000Z/DS1000Z Programming Guide
:DECoder<n>:THREshold:CHANnel1
:DECoder<n>:THREshold:CHANnel2
:DECoder<n>:THREshold:CHANnel3
:DECoder<n>:THREshold:CHANnel4
Syntax
:DECoder<n>:THREshold:CHANnel1 <thre>
:DECoder<n>:THREshold:CHANnel1?
:DECoder<n>:THREshold:CHANnel2 <thre>
:DECoder<n>:THREshold:CHANnel2?
:DECoder<n>:THREshold:CHANnel3 <thre>
:DECoder<n>:THREshold:CHANnel3?
:DECoder<n>:THREshold:CHANnel4 <thre>
:DECoder<n>:THREshold:CHANnel4?
Description
Set or query the threshold level of the specified analog channel.
Parameter
Name
Type
Range
Default
<n>
Discrete
{1|2}
--
<thre>
Real
(-4 x VerticalScale - VerticalOffset) to
(4 x VerticalScale - VerticalOffset)
--
Explanation
 VerticalScale is the vertical scale of the specified analog channel. VerticalOffset is the
vertical position of the specified analog channel.
 By default, the auto threshold function of the analog channels of the oscilloscope is
turned on. To set the threshold level manually, send
the :DECoder<n>:THREshold:AUTO command to turn off the auto threshold
function.
Return
Format
The query returns the level in scientific notation.
Example
:DECoder1:THREshold:CHANnel4 1.5 /*Set the threshold level of CH4 to 1.5V*/
:DECoder1:THREshold:CHANnel4? /*The query returns 1.500000e+00*/

Table of Contents

Other manuals for Rigol DS1104Z-S Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1104Z-S Plus and is the answer not in the manual?

Rigol DS1104Z-S Plus Specifications

General IconGeneral
BrandRigol
ModelDS1104Z-S Plus
CategoryTest Equipment
LanguageEnglish

Related product manuals