RIGOL Chapter 2 Command System
2-60 MSO1000Z/DS1000Z Programming Guide
:DECoder<n>:PARallel:NRTime
:DECoder<n>:PARallel:NRTime <time>
:DECoder<n>:PARallel:NRTime?
Set or query the noise rejection time of parallel decoding. The default unit is s.
Before sending this command, send the :DECoder<n>:PARallel:NREJect command to
turn on the noise rejection function.
The query returns the noise rejection time in scientific notation.
:DECoder1:PARallel:NRTime 0.01 /*Set the noise rejection time to 10ms*/
:DECoder1:PARallel:NRTime? /*The query returns 1.000000e-02*/
:DECoder<n>:PARallel:CCOMpensation
:DECoder<n>:PARallel:CCOMpensation <comp>
:DECoder<n>:PARallel:CCOMpensation?
Set or query the clock compensation time of parallel decoding. The default unit is s.
 Setting the compensation time can make fine adjustment of the pahse deviation
between the clock line and data line.
 This command is invalid when the CLK channel source is set to OFF
(:DECoder<n>:PARallel:CLK).
The query returns the compensation time in scientific notation.
:DECoder1:PARallel:CCOMpensation 0.01 /*Set the compensation time to 10ms*/
:DECoder1:PARallel:CCOMpensation? /*The query returns 1.000000e-02*/