EasyManuals Logo
Home>Rigol>Test Equipment>DS2000E Series

Rigol DS2000E Series User Manual

Rigol DS2000E Series
283 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #34 background imageLoading...
Page #34 background image
RIGOL Chapter 1 Quick Start
1-6 DS2000E User’s Guide
Function Inspection
1. Press Storage  Default to restore the oscilloscope to its factory state.
2. Connect the ground alligator clip of the probe to the "Ground Terminal" as
shown in the figure below.
3. Connect one end of the probe to CH1 input terminal of the oscilloscope, and
connect the other end to the "Probe Compensation Signal Output Terminal".
Figure 1-6 Probe Compensation Signal Output Terminal/Ground Terminal
4. Set the probe attenuation to 10X, and then press AUTO.
5. Observe the waveform on the display. If the waveforms actually shown do not
match that in the figure below, please perform "Probe Compensation".
Figure 1-7 Probe Compensation Signal
6. Use the same method to test the other channels.
WARNING
To avoid electric shock when using the probe, please make sure that the
insulated wire of the probe is in good condition. Do not touch the metallic
part of the probe when the probe is connected to high voltage source.
Tip
The probe compensation signal can only be used for probe compensation
adjustment and cannot be used for calibration.
Probe Compensation Signal Output Terminal
Ground Terminal

Table of Contents

Other manuals for Rigol DS2000E Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS2000E Series and is the answer not in the manual?

Rigol DS2000E Series Specifications

General IconGeneral
BrandRigol
ModelDS2000E Series
CategoryTest Equipment
LanguageEnglish

Related product manuals