9.6 Operating conditions
Location Indoor/outdoor
Altitude 5000 m (16,404 ft) maximum
Vibration resistance 0.5 g at frequencies from 10 Hz to 100 Hz
Bump/shock resistance 25 g
Ambient temperature Storage -40 to +85 °C
Operating -40 to +80 °C
Relative humidity Suitable for outdoor (TYPE 4X, TYPE 6, IP66, IP68 enclosure)
Installation category I
Pollution degree 4
EMC performance Emission EN 55011/CISPR-11
Immunity EN/IEC 61326-1 (Industry)
9.7 Process
Temperature (at flange or threads) -40 to +85 °C (-40 to +185 °F)
Pressure (vessel) 0.5 bar g/7.25 psi g
9.8 Communication
Communication type
HART
4 to 20 mA
Supported engineering systems
SIMATIC PDM
AMS Device Manager
FC375/FC475
FDT (such as Pactware, or Fieldcare)
SITRANS library for use with PCS-7
Technical data
9.8 Communication
SITRANS Probe LU240 with mA/HART
52 Compact Operating Instructions, 10/2018, A5E42673704-AB