2
2-1
Installation Verification Operator Manual 122994-513
2.1 Technical
Specifications
These specifications are intended to describe the technical infor-
mation given on the identification nameplate of the Reliance
®
400
Laboratory Glassware Washer or the Reliance
®
500 Laboratory
Glassware Washer and to state other relevant information. Verify
equipment drawing (122-992-953 for Reliance 400 or 122-993-554 for
Reliance 500) or identification nameplate (located in right bottom
corner on load side) for proper voltage and amperage.
2.1.1 Voltage, Amperage and
Power Consumption
Both laboratory glassware washers operate either on:
120/208 V, 60 Hz, three-phase, four-wire, steam-heated;
480 V, 60 Hz, three-phase, three-wire, steam-heated;
380/400/415 V, 50 Hz, three-phase, three-wires, steam-heated;
600 V, 60 Hz, three-phase, three-wire, steam-heated;
480 V, 60 Hz, three-phase, three-wire, electric-heated;
380/400/415 V, 50 Hz, three-phase, three-wire, electric-heated;
600 V, 60 Hz, three-phase, three-wire, electric-heated.
Maximum currents and power consumptions are indicated on
identification nameplate (refer to Figure 3-1).
A protective ground conductor is required (Class 1 Equipment).
Main supply voltage fluctuation not exceeding ± 10 percent of
nominal voltage.
Installation Category: Overvoltage Category II.
Overvoltage attenuation must be provided. Equipment must be
connected in a way to reduce transient voltage at electrical main
supply. Overvoltage attenuation can be achieved by:
a. connecting equipment main supply to a distribution system
with multiple branch circuits;
b. connecting equipment main supply through an isolating
transformer (380/400/415 V [input, output], three-phase,
30 kVA; 208 V [input, output], three-phase, 15 KVA; 480 V
[input, output], three-phase, 30 kVA; or 600 V [input,
output], three-phase, 30 kVA).
IMPORTANT: Customer is responsible for compliance with
applicable codes and regulations.
Refer to Uncrating/Installation Instructions (P122996-649) for proper
connection.
IMPORTANT: A listing of the Safety Precautions to be observed when operating and servicing this glass-
ware washer can be found in S
ECTION 1 of this manual. Do not operate the equipment until you have
become familiar with this information.
INSTALLATION VERIFICATION

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Steris Reliance 500 and is the answer not in the manual?

Steris Reliance 500 Specifications

General IconGeneral
BrandSteris
ModelReliance 500
CategoryWasher
LanguageEnglish

Related product manuals