5-2
122994-513 Operator Manual Cycle and Control Value Programming
PRE-WASH (1 TO 4) WASH (1 TO 5) RINSE (1 TO 4)
PURE WATER (1 TO 4)
(2)
DRYING (OPTION)
PRE-WASH 1 WASH 1 RINSE 1 RINSE 2 PURE WATER 1
PHASE: RECIRCULATED RECIRCULATED RECIRCULATED RECIRCULATED NON-RECIRCULATED RECIRCULATED
DEFAULT SELECT DEFAULT SELECT DEFAULT SELECT DEFAULT SELECT DEFAULT SELECT DEFAULT SELECT DEFAULT SELECT DEFAULT SELECT
CIRCULATION SEE
TIME
(1)
01:00 CYCLES 01:00 01:00 00:10 01:00 15:00 1:00
INJECTION PUMP 1
PUMP
SELECTION:
WATER TYPE PURE W. FROM
OR HTW
(3)
HEATED HTW HTW PURE WATER OPTIONAL TANK
AIR TEMP 65.0°C FROM TAN
K
OR
(150.0°F) (OPTION) HEATED
82°C (180°F)
MOTOR SPEED
LIGHT
LOW
(5)
X
02:00
X X X IF SELECTED X
MEDIUM
LOW
(5)
X 04:00 X X X IF SELECTED X
HEAVY
LOW
(5)
X
06:00
X X X IF SELECTED X
4 to 10
LOW
(5)
IF SELECTED IF SELECTED IF SELECTED IF SELECTED IF SELECTED IF SELECTED IF SELECTED
DESCALER
LOW
(5)
10:00 (6) 3 Rinses - 1:00 10 sec.
ADJUSTABLE IN AUTOMATIC MODE
SUMP FILL ALARM TIME: 04:00 PUMP 1 NAME: PUMP 1 INJECTION TIME (7) INJECTION UNIT CONVERSION
01:00-15:00 9 CHARACTERS
SUMP GRAVITY DRAIN ALARM TIME: 01:15 PUMP 1 00:45 MIN.:SEC. OZ/GALLON ML/L
01:00-15:00 PUMP 2 NAME: PUMP 2 00:20/03:00
SUMP HEATING ALARM: 9 CHARACTERS 00:11 1/8 1.0
•STEAM HEATED UNIT: 10:00 PUMP 2 00:45 00:22 1/4 2.0
01:00-15:00 PUMP 3 NAME: PUMP 3 00:20/03:00 00:44 1/2 3.9
•ELECTRIC HEATED UNIT: 30:00 (OPTIONAL) 9 CHARACTERS 01:06 3/4 5.9
30:00-59:00 PUMP 3 00:45 01:28 1 7.8
PUMP 4 NAME: PUMP 4 00:20/03:00 01:50 1-1/4 9.8
SUMP PUMPED DRAIN: 00:45 (OPTIONAL) 9 CHARACTERS 02:12 1-1/2 11.7
PUMP 4 00:45 02:34 1-3/4 13.7
PURE WATER TANK FILL ALARM TIME: 10:00 POWDER NAME: POWDER 00:20/03:00 02:56 2 15.6
RECOMMENDED 01:00-15:00 9 CHARACTERS
PURE WATER HEATING ALARM TIME:
NOT APPLICABLE (OPTION)
•STEAM HEATED UNIT: 10:00
01:00-15:00
VALUES ADJUSTABLE BY THE •ELECTRIC HEATED UNIT: 45:00
OPERATOR IN AUTOMATIC MODE 01:00-59:00 (1) TREATMENT TIME DOES NOT INCLUDE FILLING, HEATING, AND DRAINING TIMES.
(2) PURE WATER TREATMENT ALWAYS PERFORMED AT LOW SPEED.
CTW= COLD TAP WATER PURE WATER TANK TEMPERATURE: 82.0°C (180.0°F) (3) DEFAULT SETTING IS COLD WATER IF OPTION APPLIES.
HTW= HOT TAP WATER (OPTION) 15.0°C (60.0°F) - 88.0°C (190.0°F) (4) IF COLD WATER IS AVAILABLE.
(5) -ONE LEVEL: LOW SPEED ONLY (WITHOUT UNIVERSAL SHELVING SYSTEM)
RESERVICE IN 100 DAYS VAPOR REMOVAL TIME: -TWO LEVELS: HIGH SPEED ONLY (WITH UNIVERSAL SHELVING SYSTEM)
RESERVICE IN 1000 CYCLES •STEAM HEATED UNIT: 02:00 (6) MANUALLY ADD 25 OZ (740 mL) OF DESCALER; RATE OF 2 OZ/GAL (16 mL/L).
CYCLE COUNT: 00000001 02:00-15:00
(7) 45 SECONDS EQUAL 1/2 OZ/GAL. US (4 ml/l) APPROXIMATELY (SEE INJECTION UNIT CONVERSION).
•ELECTRIC HEATED UNIT: 02:45
REVISION 13 (2006-12-14) 01:00-15:00
REF.: #122-997-304
GENERAL NOTES:
ADJUSTABLE IN SERVICE MODE
IF SELECTED
COOLING
X
X
X
C
Y
C
L
E
S
00:15
TO
15:00
00:15
TO
15:00
00:15
TO
15:00
00:15
TO
15:00
00:15
TO
15:00
00:15
TO
30:00
CTW
(4)
HTW
CTW
(4)
HTW
CTW
(4)
HTW
HEATED
60°C
(140.0°F)
TO
88.0°C
(190.0°F)
HEATED
43.0°C
(110.0°F)
TO
88.0°C
(190.0°F)
PURE W.
HEATED
IN SUMP
15.0°C
(60.0°F) TO
88.0°C
(190.0°F)
PURE W.
FROM
TANK
65.0°C
(150°F)
TO
115.0°C
(240.0°F)
POWDER
PUMP 4
PUMP 3
PUMP 2
PUMP 1
NONE
X
AIR HEATED AT
115.0°C (240.0°F)
SET POINT, NOT
MAINTAINED
HEATED
43.0°C
(110.0°F)
TO
88.0°C
(190.0°F)
00:00
TO
30:00
Table 5-1. Reliance 400 / Reliance 500 Laboratory Glassware Washer – Cycle Description Chart